Giovanni, 01.08.2024
LaBr3_900V_UG_Am_20240801_094057.root → 1 min, threshold too high, visible some gamma and alpha peak?LaBr3_900V_UG_Th_20240801_094933.root → 1 min, 10 RMS threshold, visible several peaks. I suspect the upper region are mostly alphas (different Chi2)LaBr3_900V_UG_Am_20240801_101756.root → 1 min, 5 RMS threshold, the 59 keV peak from Am should be the one on the very left, before the continuum.LaBr3_900V_UG_bkg_20240801_102059.root → 1 min, 5 RMS threshold. The peak on the very left is not visible anymore. Very large peak in the middle of the spectrum, with a resolution of 1.16% (sigma). Maybe 40K?LaBr3_950V_UG_bkg_20240801_102506.root → 1 min, 5 RMS threshold. Resolution of the unknown peak: 1.04% (sigma).LaBr3_950V_UG_Am_20240801_103450.root → 1 min, 5 RMS threshold. 59 keV peak from 241Am visible on the left of the spectrum. At this point the unknown peak could be the 662keV peak from 137Cs, even though I have the suspicion the response is non-linear.LaBr3_950V_UG_Th_20240801_104540.root and LaBr3_950V_UG_Th_20240801_105223.root –> 1 min, first run with “unshielded” source, second run with source embedded in few mm thick lead layer. Unclear how to interpret the spectrum.LaBr3_1000V_UG_bkg_20240801_105721.root → 1 min, 5 RMS threshold. Resolution of unknown peak: 0.99% (sigma).LaBr3_1000V_UG_Am_20240801_110337.root → 1 min, 5 RMS threshold. 59 keV gamma from 241Am clearly visible.LaBr3_1000V_UG_Th_20240801_111101.root → 10 min, 5 RMS threshold.LaBr3_1050V_UG_bkg_20240801_112617.root → 1 min, 5 RMS threshold. Resolution on unknown peak: 0.84% (sigma)LaBr3_1100V_UG_bkg_20240801_113554.root → 1 min, 5 RMS threshold. Resolution on unknown peak: 0.81% (sigma)LaBr3_1150V_UG_bkg_20240801_113948.root → 1 min, 5 RMS threshold. Resolution on unknown peak: 0.81% (sigma)LaBr3_1200V_UG_bkg_20240801_114335.root → 1 min, 5 RMS threshold. Resolution on unknown peak: 0.77% (sigma)Since the PMT seems to saturate even at 900 V, let's do another HV scan down to lower voltage.
LaBr3_900V_UG_Th_20240820_153224.root –> 2 min, 10 RMS threshold.LaBr3_850V_UG_Th_20240820_155012.root –> 2 min, 10 RMS threshold.Detector re-assembled to try making a precision measurement of the 215Po half-life using the delayed coincidence technique.
Detector parameters:
Acquired data:
LaBr3_H2431_1250V_UG_232Th_20241126_170543.rootLaBr3_H2431_1250V_UG_bkg_20241126_180924.root to …Processed data:
/data/Reconstructed_LaBr3/global→Draw(“chi2_poweraverage.amplitude/integral:integral»(1000,,,1000)”,“fabs(slope)<1.e8 && delay>8.5e-8 && delay<1.15e-7”,“colz”)