Dedicated measurements for 154Gd decay
1) First measurement with Pb only
data taken from
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260312_100728.root to
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260316_142013.root
2) Second measurement
Moved the setup to SABRE which has a better shielding (15 cm Pb + 5 cm Cu + N2 flushing), and the data are taken from
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260316_163612.root to
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260317_141213.root
(note: /data/GAGG_1.4kV_UG_bkg_AmpX10filtered_20260316_161411.root is the file with the amplifier connected to the first output)
3) Following the second measurement, the stainless steel holder is removed
This is to understand where the internal contamination comes from
Data are taken from
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260317_142726.root to
GAGG_1.4kV_UG_bkg_AmpX10filtered_20260318_134819.root
4)
Measurement with low background gamma spectrometry (Crystal and PMT) to understand the source of the background
~590 keV bump and Potassium
5) After confirmation that 40K is mainly from the Borosilicate glass window of the PMT, a new measurement campaign has started with Giuseppe's PMT which has a quartz glass, which doesn't have so much 40K contamination inside
Processed data
Each data file lasts ten minutes, and for the measurement with the Pb shielding measured in gator,
cd /data/Reconstructed_GAGG_Pb/
The data taken with Pb + Cu + N2 flushing in the SABRE experiment hub
cd /data/Reconstructed_GAGG_sabre/
After removing the stainless steel holder, and continued the measurement in the SABRE
cd /data/Reconstructed_GAGG_sabre_no_steel_holder/